“…The underestimation of the electron saturation current and/or the overestimation of the ion saturation current are the likely causes of this decrease in the SCR. The reasons for the underestimation of the electron saturation current are generally attributed to several factors, such as, probe surface contamination [54,55], an increase in plasma ground sheath resistance [55,56], the presence of an external magnetic field [55,[57][58][59][60], and non-Maxwellian [19,55,61,62] or drifting plasmas [63]. This leads to an underestimation of the plasma potential, either determined from the knee in the amperevoltage characteristic or using the first derivative method [56].…”