Uniquely shaped cardiovascular stents enhance the pressure generation of intravascular blood pumps

Abstract: The specialized design of the stents, which protect the vessel wall from the rotating components of the pump, proved to be advantageous by further augmenting the pressure generation of the pump, reducing the flow vorticity at the pump outlet, and enhancing flow control.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1

Citation Types

0
19
0

Year Published

2014
2014
2023
2023

Publication Types

Select...
6

Relationship

3
3

Authors

Journals

citations
Cited by 9 publications
(19 citation statements)
references
References 15 publications
0
19
0
Order By: Relevance
“…Figure illustrates the stent designs as investigated. The super‐diffuser has straight filaments with a set of diffuser blades on the stent at the outlet, whereas the against–with cage has filaments partially twisted in the opposite direction to the impeller blades and filaments partially twisted in the same direction as the diffuser blades . The two cages are the result of two separate design iterations incorporating hydraulic testing .…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…Figure illustrates the stent designs as investigated. The super‐diffuser has straight filaments with a set of diffuser blades on the stent at the outlet, whereas the against–with cage has filaments partially twisted in the opposite direction to the impeller blades and filaments partially twisted in the same direction as the diffuser blades . The two cages are the result of two separate design iterations incorporating hydraulic testing .…”
Section: Methodsmentioning
confidence: 99%
“…These surgical treatment strategies are limited in long-term benefit, and few therapeutic alternatives exist. The lack of therapeutic options for Fontan patients necessitates the development of new cavopulmonary assist doi: 10.1111/aor.12255 devices to address the need for alternative circulatory support for this growing cohort of patients (11)(12)(13)(14)(15)(16)(17). A mechanical blood pump specifically designed to increase pressure in the great veins would improve hemodynamic stability in adolescent and adult patients having dysfunctional cavopulmonary circulations.…”
mentioning
confidence: 99%
“…The performance achievements met the desired pressure-flow requirements to support a Fontan patient (3,(5)(6)(7)(8). All of the pump designs produced a pressure rise over the flow range evaluated and rotational speeds of 4000-7000 RPM.…”
Section: Discussionmentioning
confidence: 76%
“…Details about this flow loop and the methodology that was employed can be found in the literature (5,7). A shaft-driven configuration was utilized for testing.…”
Section: Experimental Testing Of Impeller Prototypementioning
confidence: 99%