A 17-year-old male with hypertrophic cardiomyopathy underwent placement of a dual-chamber implantable cardioverter defibrillator (ICD) in 2010. In October 2016, he suffered a cardiac arrest. His ICD was interrogated after visiting our hospital. The shock failed, and ventricular fibrillation (VF) terminated spontaneously to sinus rhythm. The Durata lead (SJM, Sylmar, CA, USA) was removed for investigation. The lead had an internal abrasion of a right ventricular conductor cable at the superior vena cava coil. These findings suggest that the ICD shocks that failed to terminate the VF were delivered with low energy due to an internal short circuit of the Durata lead. The very short charge time indicates insufficient energy delivery, which may cause failure of cardioversion and defibrillation.