2007
DOI: 10.5435/00124635-200701000-00006
|View full text |Cite
|
Sign up to set email alerts
|

Wear and Osteolysis Around Total Knee Arthroplasty

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

1
129
0
2

Year Published

2008
2008
2023
2023

Publication Types

Select...
6
2

Relationship

0
8

Authors

Journals

citations
Cited by 204 publications
(134 citation statements)
references
References 67 publications
1
129
0
2
Order By: Relevance
“…With the contact stresses transmitted during physiological loading, the severity of articular wear is most influenced by the quality of the polyethylene, level of implant constraint, insert thickness, and coronal tibiofemoral alignment [4,26,30,36]. The axial rolling, sliding, and rotation forces generate articular-sided polyethylene wear debris that is relatively larger in size as a byproduct of subsurface cracking, delamination, and pitting [4,30].…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…With the contact stresses transmitted during physiological loading, the severity of articular wear is most influenced by the quality of the polyethylene, level of implant constraint, insert thickness, and coronal tibiofemoral alignment [4,26,30,36]. The axial rolling, sliding, and rotation forces generate articular-sided polyethylene wear debris that is relatively larger in size as a byproduct of subsurface cracking, delamination, and pitting [4,30].…”
Section: Discussionmentioning
confidence: 99%
“…Several factors have been identified that influence backside wear, including insert locking mechanism, tibial baseplate surface, implant duration, patient activity level, and tibiofemoral alignment [6,7,26,29,34]. Modular insert locking designs may use a mechanism of full peripheral capture, partial capture, full tongue-and-groove, or partial tongue-and-groove.…”
Section: Discussionmentioning
confidence: 99%
“…We understand this precludes any conclusions regarding long-term differences in the weight, size, and shape of polyethylene wear particles in the TKAs with an OxZr femoral component and in those with a Co-Cr femoral component. Although polyethylene wear particles from failed knee prostheses yield important information, polyethylene wear particles in such failed TKA prostheses are generated in extreme conditions [2,23] and might be affected by unstable interface and third bodies; thus we believe such data from patients who are functioning well yield complementary data.…”
Section: Discussionmentioning
confidence: 99%
“…Tibial component polyethylene wear, with resultant particle-induced osteolysis, is one of the causes of late failure, component loosening, and reoperation after TKA [13,21,23]. The frequency of periprosthetic osteolysis after TKA has been reported to be 5% to 20% at followup of \ 5 to 15 years [1,6,15,16,21].…”
Section: Introductionmentioning
confidence: 99%
“…The frequency of periprosthetic osteolysis after TKA has been reported to be 5% to 20% at followup of \ 5 to 15 years [1,6,15,16,21]. The prevalence and severity of polyethylene wear after TKA are related to multiple factors, including implant design, patient age and activity level, and technical issues.…”
Section: Introductionmentioning
confidence: 99%