There are three principal magnetic fields in magnetic resonance imaging (MRI) that may interact with medical implants. The static field will induce force and torque on ferromagnetic objects. The pulsed gradients are of audio frequency and the implant may concentrate the induced currents, with a potential for nerve stimulation or electrical inference. The currents induced in the body by the radio frequency (RF) field may also be concentrated by an implant, resulting in potentially dangerous heating of surrounding tissues. This paper presents basic information about MRI interactions with implants with an emphasis on RF-induced heating of leads used for deep brain stimulation (DBS). The temperature rise at the electrodes was measured in vitro as a function of the overall length of a DBS lead at an RF frequency of 64 MHz. The maximal temperature rise occurred for an overall length of 41 cm. The method of moments was used to calculate the current induced in the lead. From the induced currents, the RF power deposition near the electrodes was calculated and the heat equation was used to model the temperature rise. The calculated temperature rises as a function of lead length were in good agreement with the measured values.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.