We present a case of a patient submitted for extirpation of a neoplasm of the larynx, by means of carbon dioxide laser surgery. High frequency jet ventilation was applied by means of orotracheal intubation with two Teflon catheters, 2 mm in external diameter and 30 cm in length, attached with three equally placed strips of adhesive paper tape. One catheter was used to inject the jet volume and the other used to measure the airway pressure. The adhesive strips were moistened and FiO2 was lower than 50%. After 30 min using the laser, an airway fire was noticed. Ventilation was interrupted and the catheters were removed. The patient was reintubated with an endotracheal tube of 6 mm ID and the surgical procedure was continued until the tumour was removed. Two factors contributed to the airway fire: the ignition of the lowest adhesive strip that had dried and the use of the laser in the mode of continuous pulsation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.