To determine the optimal collimation, pitch and reconstruction interval for CT colonography, 10 spherical polyps between 1 mm and 10 mm diameter and made of tissue equivalent material with a CT number of 40 Hounsfield units (HU) were placed in the colon of an anthropomorphic phantom. The phantom was scanned at slice thicknesses of 3 mm, 5 mm and 7 mm and pitches of 1.0, 1.3, 1.5, 1.7 and 2.0 on an IGE Hispeed advantage system. Images were reconstructed for each scanning parameter at the minimum intervals allowed along the z-axis. The optimum scanning protocol was assessed by measuring maximum contrast between the polyp and air, sensitivity for detection of each polyp along the z-axis, and relative radiation dose. In addition, images were reviewed separately by two radiologists who graded polyp conspicuity as: 0, not seen; 1, faintly seen; 2, well seen. It was found that varying the scanning parameters caused a marked alteration in the maximum contrast between each polyp and air. For example, for the 5 mm polyp, the range of contrasts from best to worst case was 910-490 HU. It was noted that with contrasts of less than 500 HU, polyps were only faintly seen. A slice thickness of 3 mm with a pitch of 2 offers optimal polyp conspicuity with a relatively low radiation dose, we conclude that scanning parameters can be optimized for threshold contrast, radiation dose and subjective conspicuity. We propose an optimal parameter of 3 mm slice thickness and pitch 2.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.