The dose-response of LiF:Mg,Ti (TLD-100) exposed to 15 and 35 kVp (8.0 +/- 0.1 and 8.1 +/- 0.1 keV effective energy respectively) x-rays and 60Co gamma-rays has been measured in the dose interval from (1.2-5.4) x 10(3) Gy for x-rays, and from 0.14 to 850 Gy for gamma-rays. In both cases the total TL signal and glow curve peaks 3 to 9 show supralinearity. The supralinearity function f(D) is similar for both x-ray beams, except for peak 8, where a 30% difference is observed. The maxima of f(D) for the total TL signal and peaks 5 to 8 are 2.1, 1.7, 6.4, 3.3 and 7.5 respectively for 8.1 keV x-rays and 3.7, 3.1, 13.6, 9.9 and 11.0 for gamma-rays. The measured relative efficiencies for x-rays with respect to 60Co, for the total TL signal and peaks 5 and 7, were 1.04, 0.97 and 3.2 respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.