Purpose:To evaluate plan quality of a novel MRI-compatible direction modulated brachytherapy (DMBT) tandem applicator using 192 Ir, 60 Co, and 169 Yb HDR brachytherapy sources, for various cervical cancer high-risk clinical target volumes (CTV HR ). Materials and Methods: The novel DMBT tandem applicator has six peripheral grooves of 1.3-mm diameter along a 5.4-mm thick nonmagnetic tungsten alloy rod. Monte Carlo (MC) simulations were used to benchmark the dosimetric parameters of the 192 Ir, 60 Co, and 169 Yb HDR sources in a water phantom against the literature data. 45 clinical cases that were treated using conventional tandemand-ring applicators with 192 Ir source ( 192 Ir-T&R) were selected consecutively from intErnational MRI-guided BRAchytherapy in CErvical cancer (EMBRACE) trial. Then, for each clinical case, 3D dose distribution of each source inside the DMBT and conventional applicators were calculated and imported onto an in-house developed inverse planning optimization code to generate optimal plans. All plans generated by the DMBT tandem-and-ring (DMBT T&R) from all three sources were compared to the respective 192 Ir-T&R plans. For consistency, all plans were normalized to the same CTV HR D90 achieved in clinical plans. The D 2 cm3 for organs at risk (OAR) such as bladder, rectum, and sigmoid, and D90, D98, D10, V100, and V200 for CTV HR were calculated. Results: In general, plan quality significantly improved when a conventional tandem (Con.T) is replaced with the DMBT tandem. The target coverage metrics were similar across 192 Ir-T&R and DMBT T&R plans with all three sources (P > 0.093). Yb source generally resulted in the greatest OAR sparing when the CTV HR were larger and irregular in shape, while for smaller and regularly shaped CTV HR (<30 cm 3 ), OAR sparing between the sources were comparable. Conclusions: The DMBT tandem provides a promising alternative to the Con.T design with significant improvement in the plan quality for various target volumes. The DMBT T&R plans generated with the three sources of varying energies generated superior plans compared to the conventional T&R applicators. Plans generated with the 169 Yb-DMBT T&R produced best results for larger and irregularly shaped CTV HR in terms of OAR sparing. Thus, this study suggests that the combination of the DMBT tandem applicator with varying energy sources can work synergistically to generate improved plans for cervical cancer brachytherapy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.