Aims:The aim of this study was to assess the efficacy and safety of the Occlutech Paravalvular Leak Device (PLD) for the percutaneous closure of paravalvular leaks (PVL).Methods and results: Patients with PVL were enrolled at 21 sites from nine countries. Indications for PVL closure were heart failure and/or haemolytic anaemia. Endpoint measures were changes in PV regurgitation grade, NYHA class and requirement for haemolysis-related transfusion. One-hundred and thirty-six patients with mitral (67.6%) or aortic (32.4%) leaks were included (mean age 66.7 years, 58% male); 31% had multiple PVLs. The proportion of patients with NYHA Class III/IV decreased from 77.3% at baseline to 16.9% at latest follow-up. The proportion of patients with need for haemolysis-related blood transfusion decreased from 36.8% to 5.9% and from 8.3% to 0% for ML patients and AL patients, respectively. Allcause mortality was 7.4%. Complications included interference with valve leaflets (0.7%), transient device embolisation (percutaneously solved) (0.7%), late device embolisation (0.7%), recurrent haemolytic anaemia (2.2%), new-onset haemolytic anaemia (0.7%), valve surgery (2.2%), need for repeat closure (0.7%), complications at femoral puncture site (0.7%) and arrhythmias requiring treatment (4.4%).Conclusions: PVL closure with the Occlutech PLD demonstrated a high success rate associated with significant clinical improvement and a relatively low rate of serious complications.
Between December 2014 and March 2021, 144 patients with aortic (Ao) or mitral (Mi) paravalvular leaks (PVLs) were enrolled at 21 sites in 10 countries. Safety data were available for 137 patients, who were included in the safety analysis fraction (SAF), 93 patients with Mi PVLs and 44 patients with Ao PVLs. The full analysis set (FAS) comprised 112 patients with available stratum (aortic/mitral leak) as well as baseline (BL), 180-day or later assessments (2 years). Procedural success (implantation of the device with a proper closure of the PVL, defined as reduction in paravalvular regurgitation of ≥one grade as assessed by echocardiography post implantation) was achieved in 91.3% of FAS patients with Mi PVLs and in 90.0% of those with Ao PVLs. The proportion of patients suffering from significant or severe heart failure (HF), classified as New York Heart Association (NYHA) class III/IV, decreased from 80% at baseline to 14.1% at 2-year follow-up (FAS). The proportion of FAS patients needing hemolysis-related blood transfusion decreased from 35.5% to 3.8% and from 8.1% to 0% in Mi patients and Ao patients, respectively. In total, 35 serious adverse events (SAEs) were reported in 27 patients (19.7%) of the SAF population. The SAEs considered possibly or probably related to the device included device embolization (three patients), residual leak (two patients) and vascular complication (one patient). During follow-up, 12/137 (8.8%) patients died, but none of the deaths was considered to be device-related. Patients implanted with the Occlutech Paravalvular Leak Device (PLD) showed long-lasting improvements in clinical parameters, including NYHA class and a reduced dependency on hemolysis-related blood transfusions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.