The generation of synthetic CT for carbon ion radiotherapy (CIRT) applications is challenging, since high accuracy is required in treatment planning and delivery, especially in an anatomical site as complex as the abdomen. Thirty-nine abdominal MRI-CT volume pairs were collected and a three-channel cGAN (accounting for air, bones, soft tissues) was used to generate sCTs. The network was tested on five held-out MRI volumes for two scenarios: (i) a CT-based segmentation of the MRI channels, to assess the quality of sCTs and (ii) an MRI manual segmentation, to simulate an MRI-only treatment scenario. The sCTs were evaluated by means of similarity metrics (e.g., mean absolute error, MAE) and geometrical criteria (e.g., dice coefficient). Recalculated CIRT plans were evaluated through dose volume histogram, gamma analysis and range shift analysis. The CT-based test set presented optimal MAE on bones (86.03 ± 10.76 HU), soft tissues (55.39 ± 3.41 HU) and air (54.42 ± 11.48 HU). Higher values were obtained from the MRI-only test set (MAEBONE = 154.87 ± 22.90 HU). The global gamma pass rate reached 94.88 ± 4.9% with 3%/3 mm, while the range shift reached a median (IQR) of 0.98 (3.64) mm. The three-channel cGAN can generate acceptable abdominal sCTs and allow for CIRT dose recalculations comparable to the clinical plans.
Objective: Respiration negatively affects the outcome of a radiation therapy treatment, with potentially severe effects especially in particle therapy (PT). If compensation strategies are not applied, accuracy cannot be achieved. To support the clinical practice based on 4D Computed Tomography (CT), 4D Magnetic Resonance Imaging (MRI) acquisitions can be exploited. The purpose of this study was to validate a method for virtual 4DCT generation from 4DMRI data for lung cancers on a porcine lung phantom, and to apply it to lung cancer patients in PT.
Approach: Deformable image registration was used to register each respiratory phase of the 4DMRI to a reference phase. Then, a static 3DCT was registered to this reference MR image set, and the virtual 4DCT was generated by warping the registered CT according to previously obtained deformation fields. The method was validated on a physical phantom for which a ground truth 4DCT was available and tested on lung tumor patients, treated with gated PT at end-exhale, by comparing the virtual 4DCT with a re-evaluation 4DCT. The geometric and dosimetric evaluation was performed for both proton and carbon ion treatment plans.
Main results: The phantom validation exhibited a geometrical accuracy within the maximum resolution of the MRI and mean dose deviations, with respect to the prescription dose, up to 3.2% for target D95%, with a mean gamma pass rate of 98%. For patients, the virtual and re-evaluation 4DCTs showed good correspondence, with errors on target D95% up to 2% within the gating window. For one patient, dose variations up to 10% at end-exhale were observed due to relevant inter-fraction anatomo-pathological changes that occurred between the planning and re-evaluation CTs. 
Significance: Results obtained on phantom data showed that the virtual 4DCT method was accurate, allowing its application on patient data for testing within a clinical scenario.

scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.