The aim of this study was to estimate the radioactive risk for surgical staff performing radioguided sentinel lymph node (SN) biopsy and to calculate the contamination level in the operating room for assessment of the possible need for specific radiation protection procedures. We studied 20 patients who were selected for quadrantectomy and SN biopsy. The day before surgery a volume of 0.15 mL of 99mTc-nanocoll was injected: the activity was 3.11 +/- 0.85 MBq in group A (15 pts) and 11.6 +/- 0.6 MBq in group B (5 pts). External radiation to staff was evaluated by measuring the exposure rate in air one hour after radiopharmaceutical administration. The air KERMA rate during surgery was estimated considering the physical decay of 99mTc. Contamination of disposable materials and surgical instruments in the operating room was measured using a contamination monitor, whereas the residual activity in the SN and the injection site was measured with a gamma probe. The exposure rate at 20 cm from the injection site was 0.75 microSv/h when the most radioactive patients (group B) were treated. Contamination in the operating room proved to be negligible. Considering the number of radioguided treatments carried out by a surgeon in one year, an equivalent effective dose of 0.075 mSv was estimated; the recommended dose limit according to the relevant Italian law, DL 230/95, is 1 mSv/yr. Surgical staff therefore do not require a classification of "exposed workers" and there is no need to supply the operating room with special containers for radioactive waste.
Background:
An accurate measurement of the target volume is of primary importance in theragnostics of hyperthyroidism
Objective:
Our purpose was to evaluate the accuracy of a threshold–based isocontour extraction procedure for thyroid tissue volumetry from SPECT-CT.
Methods:
Cylindrical vials with a fixed volume of 99mTcO4 at different activities were inserted into a neck phantom in two different thickness settings. Images were acquired by orienting the phantom in different positions, i.e., 40 planar images and 40 SPECT-CT. The fixed values of the iso-contouring threshold for SPECT and SPECT-CT were calculated by means of linear and spline regression models. Mean, Median, Standard Deviation, Standard Error, Mean Absolute Percentage Error and Root Mean-Square Error were computed. Any difference between the planar method, SPECT and SPECT-CT and the effective volume was evaluated by means of ANOVA and post-hoc tests. Moreover, planar and SPECT-CT acquisitions were performed in 8 patients with hyperthyroidism, considering relevant percentage differences greater than > 20 % from CT gold standard.
Results:
Concerning phantom studies, the planar method shows higher values of each parameter than the other two methods. SPECT-CT shows lower variability. However, no significant differences were observed between SPECT and SPECT-CT measurements. In patients, relevant differences were found in 7 out of 9 lesions with the planar method, in 6 lesions with SPECT, but in only one with SPECT-CT.
Conclution:
Our study confirms the superiority of SPECT in volume measurement if compared with the planar method. A more accurate measurement can be obtained from SPECT-CT.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.