A validated method was developed for determination of diclofenac sodium, considered a model hydrophilic drug for in vitro permeation studies, in Eagle's Minimum Essential Medium (MEM) with Earle's balanced salt solution. Liquid chromatography was used to determine diclofenac sodium. This method was developed with a reversed-phase column Supercosil LC 18, DB 25 cm × 4.5 mm; the mobile phase was methanol with 3% (v/v) acetic acid–Milli-Q water (74 + 26), and detection was at 283 nm. The detection and quantitation limits were 2.41 × 10–8 and 3.31 × 10–5 μg/μL, respectively. The accuracy within-day (n = 3) and day-to-day (n = 7) was 98.83%; the mean variation coefficient for inter- (n = 7) and intraday precision (n = 3) was 12.20%, thus, not exceeding 15%. This method can be used as an analytical procedure for the determination of diclofenac sodium in MEM for in vitro permeation studies.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.