A selective and sensitive high-performance liquid chromatography method with fluorescence detection for simultaneous determination of irbesartan, losartan and valsartan on dried blood spots (DBS) has been developed and validated. It involves solvent extraction of a punch of DBS followed by reversed-phase liquid chromatography on a Lichrospher(®) 100 RP-18e column. Fluorescence detection was performed at 259 and 385 nm as excitation and emission wavelengths, respectively. The detection limits of irbesartan, losartan and valsartan were 1.8, 3.6 and 1.8 ng/mL respectively. The mean recoveries of irbesartan, losartan and valsartan were 98.68, 98.42 and 97.81%, respectively. The mean inter-day and intra-day precisions of irbesartan, losartan and valsartan were 2.07 and 1.34%, 1.42 and 1.48%, and 3.20 and 2.15% respectively. The proposed method was simple and rapid. Design of experiments was used to evaluate the robustness of the method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.