[1] Element concentration and isotope ratio measurements by single-collector mass spectrometry often require the detection system to handle ion beams with very large intensity ratios. In order to obtain accurate and reproducible element concentration and isotope ratio data, the detection system must have a linear response with respect to the intensity of the incident ion beam. An extended range scaling pulse counting detector equipped on a Varian 810 quadrupole inductively coupled plasma-mass spectrometer (ICP-MS) was tested for linearity across count rates of ∼2000 to 110,000,000 cps with different concentrations of natural U solutions. We also tested detector linearity by the laser ablation analysis of 206 Pb/ 238 U, 207 Pb/ 235 U, and 207 Pb/ 206 Pb ratios in well-characterized 416-1565 Ma zircon standards. Results indicate that there is no correlation between the measured isotope ratio and ion intensity for the solution tests or the tests of natural zircon standards. The results of these tests confirm the suitability of this instrument for isotope ratio measurements that require a substantial dynamic range without having to switch between pulse counting and analog modes on electron multipliers or switching between electron multiplier to Faraday detectors.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.