Summary:The selective multitest Boehringer Mannheim/Hitachi 704 analysis System was examined according to the ECCLS guidelines in a multicentre evaluation involving four laboratories. Ten routine parameters, covering most of the application settings of the Instrument, were measured in the respective laboratory at temperatures 25, 30 or 37 °C. The trial lasted four months and gave more than 40000 data. It yielded the following results:1. Within the four laboratories the mean coeffidents of Variation for three control sera at different concentrations were found to be equal to or better than 1.6% for the within-run imprecision and 2.8% or better for the between-day imprecision.2. No drift was observed during eight hours.3. Because of the high linear measuring ränge a re-run analysis was seldom necessary.4. 'Sample-related carry-over was not seen. Reagent-dependent carry-over was measured from cholesterol to uric acid and from triacylglycerols to lipase. Through modification of the cholesterol and triacylglycerol reagents, the carry-over effect was präctically eliminated.5. The recovery of the assigned values of control sera showed average values between 99 and 104%. For bilirübin, creatinine, creatine kinase and alanine aminotransferase some control sera showed deviations greater than 10%.6. In all cases, regression analysis of the resülts obtained in comparisons of the present Instrument with the Hitachi 705 or 737 yielded slopes close to unity with extreme values of 0.95 and 1.06.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.