Summary:The ES 600 sample^selective multibatch analyser was subjected to a multicentre evaluation in six laboratories in accordance with ECCLS guide-lines. During the 3-month trial, five Enzymun-Test® diagnostics (T 4 , TBG, Digoxin, CEA and TSH)') were measured at 25 °C.The study yielded the following results:1. The within-series and between-series precision were very good, with mean CV's of approx. 3% and 7% respectively.2. Recovery of the target values for three control sera was in the range ±5%. 5. Within the measuring range defined by the standards, no deviations could be ascertained upon dilution of the samples.6. Total carry-over in the instrument was below 0.05%.From studies with instruments from the first production series, it became evident that modifications were necessary to improve the reliability. A follow-up measuring programme confirmed a clear improvement in reliability and a reduction in the imprecision, particularly for results from series to series.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.