Silicon photomultipliers (SiPMs) are an emerging solid-state alternative to photomultiplier tubes (PMTs) for low light detection, with similar gain but lower cost and lower operating voltage. We demonstrate coherent anti-Stokes Raman scattering (CARS) imaging in a side-by-side comparison of an uncooled SiPM with an uncooled multialkali PMT as well as a state-of-the-art cooled GaAsP PMT. We determine the optimum reverse-bias voltage for acquiring the best signal-to-noise ratio (SNR) for CARS imaging of lipids at 2850 c m − 1 . We find that despite the higher dark counts, the SNR of CARS images acquired with the uncooled SiPM biased at an optimum voltage is better than that of the multialkali PMT and close to that of the cooled GaAsP PMT ( ∼ 1.5 and ∼ 0.8 times, respectively). This is due to the higher gain and lower excess noise factor related to the pulse height variability in the SiPM.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.