The analytical characteristics of the AxSYM Ultrasensitive hTSH-II (Abbott Laboratories) procedure for quantitation of serum thyrotropin (TSH) concentration were evaluated. Within- and between-run imprecisions, functional sensitivity, analytical interval and relative inaccuracy with respect to the Enzymum-Test TSH (Boehringer Mannheim) were studied. In all cases, the within-run and between-run of coefficients variation were lower than 6.69% and 8.12% respectively. The measurement range was tested with serial dilutions of a serum with a high thyrotropin concentration, and the procedure was found to be linear up to at least 87.0 mIU/l. The functional sensitivity was 0.018 mIU/l. The relative inaccuracy study (Passing-Bablok non-parametric linear regression) produced the following linear equation: (AxSYM) = 1.02. (ES-700)-0.03 mIU/l, with 95% confidence intervals of a (-0.05; -0.01); b (0.98; 1.06).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.