Within the framework of the JRP06 European project ‘Increasing Cancer Treatment Efficacy Using 3D Brachytherapy’, a prototype of a graphite standard calorimeter for the measurement of the absorbed dose rate to water,
, for 192Ir sources used in high dose rate (HDR) brachytherapy has been developed at the Italian National Institute of Ionizing Radiation Metrology (ENEA-INMRI). The calorimeter was tested at the Sant'Andrea Hospital in Rome, where
measurements were performed in the quasi-adiabatic mode of operation using an 192Ir MicroSelectron® HDR V2 source. The
measurements showed a reproducibility of about 1%, while the combined standard uncertainty on the
value at the distance of 1 cm from the source was estimated as 1.4%, lower than the uncertainty of
determined from the reference air-kerma rate.
BackgroundCure rate of early Hodgkin Lymphoma are high and avoidance of late toxicities is of paramount importance. This comparative study aims to assess the normal tissue sparing capability of intensity-modulated radiation therapy (IMRT) versus standard three-dimensional conformal radiotherapy (3D-CRT) in terms of dose-volume parameters and normal tissue complication probability (NTCP) for different organs at risk in supradiaphragmatic Hodgkin Lymphoma (HL) patients.MethodsTen HL patients were actually treated with 3D-CRT and all treatments were then re-planned with IMRT. Dose-volume parameters for thyroid, oesophagus, heart, coronary arteries, lung, spinal cord and breast were evaluated. Dose-volume histograms generated by TPS were analyzed to predict the NTCP for the considered organs at risk, according to different endpoints.ResultsRegarding dose-volume parameters no statistically significant differences were recorded for heart and origin of coronary arteries. We recorded statistically significant lower V30 with IMRT for oesophagus (6.42 vs 0.33, p = 0.02) and lungs (4.7 vs 0.1 p = 0.014 for the left lung and 2.59 vs 0.1 p = 0.017 for the right lung) and lower V20 for spinal cord (17.8 vs 7.2 p = 0.02). Moreover the maximum dose to the spinal cord was lower with IMRT (30.2 vs 19.9, p <0.001). Higher V10 with IMRT for thyroid (64.8 vs 95, p = 0.0019) and V5 for lungs (30.3 vs 44.8, p = 0.03, for right lung and 28.9 vs 48.1, p = 0.001 for left lung) were found, respectively. Higher V5 and V10 for breasts were found with IMRT (V5: 4.14 vs 20.6, p = 0.018 for left breast and 3.3 vs 17, p = 0.059 for right breast; V10: 2.5 vs 13.6 p = 0.035 for left breast and 1.7 vs 11, p = 0.07 for the right breast.) As for the NTCP, our data point out that IMRT is not always likely to significantly increase the NTCP to OARs.ConclusionsIn HL male patients IMRT seems feasible and accurate while for women HL patients IMRT should be used with caution.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.