In order to extend the French water flow calibration range, LNE-CETIAT is on progress to build a new reference for flow rate measurements ranging from 10 000 ml.h -1 down to 1 ml.h -1 . This new facility based on a gravimetric principle will enable calibration with water temperature between 10 °C and 50 °C. Flow is regulated by the combined use of a tightly regulated "upstream" pressure (0,1 to 10 bar) and the use of capillaries under laminar flow. The expected relative uncertainty for the water flow rate is 0.1 % with a coverage factor k = 2 (95% confidence). The paper described the design and the ongoing implementation of this new calibration facility.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.