Modular neck femoral stems have been associated with adverse local tissue reactions (ALTR), leading to a voluntary recall, but these effects have not been well-characterized. A retrospective review of intraoperative findings and cobalt/chromium levels was performed in 103 hips undergoing revision for ALTR. The average preoperative serum cobalt level was 7.6 μg/L (range 1.1-23 μg/L) and chromium level was 1.8 μg/L (range 0.1-6.8 μg/L). Metallic sludge was noted in 100%, synovitis in 98%, pericapsular rind in 82%, and calcar erosion in 85%. An osteotomy was required for removal in 44%. We concluded that revision of modular neck femoral stems is associated with increased preoperative metal ion levels and stem-neck corrosion. Despite advanced stem explantation techniques, osteotomy was frequently required, leading to increased morbidity.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.