Introduction
Given the fragility of reusable ureterorenoscopes, many single use instruments have appeared on the market. Unfortunately, reuse of these scopes occurs in undeveloped countries in order to cut costs. This raises safety concerns for the patient.
The aim of this article was to macroscopically evaluate the changes that single use flexible ureterorenoscopes (su-fURS) suffer after a retrograde intrarenal surgery (RIRS), and to compare them to reusable fURS.
Material and methods
Pre and post-operative images of the instruments used in 23 RIRS were obtained. All the cases had renal calculi of the inferior calix between 10–15 mm, and all of them were treated with Holmium laser. The ureterorenoscopes used were: Storz
®
Flex X2, Storz
®
Flex XC, Pusen
®
3022, OTU
®
Wiscope, AnQIng
®
Innovex and Boston Scientific
®
LithoVue. Qualitative comparisons of these were made.
Results
After su-fURS usage, significant damage was observed, especially on the distal tip. Deflection was not compromised. Reusable fURS did not sustain any damage after their use.
Conclusions
fURS are delicate equipment, especially if they are of single use. The considerable damage sustained by single use scopes could mean that reuse of these instruments is dangerous and should be avoided.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.