The skull causes topographical mislocalization in range of images of the brain due to the increased velocity of sound through its varying thickness. It also causes mislocalization in azimuth by refraction. Similarly, mislocalization in azimuth may result from echoes off the central axis of the transducer but distant to sonolucent areas of the skull, being displayed as if they lay in the central axis. Resolving power is degraded in range by the display of ringing due to either increased amplitude from strong echoes or reinforcing reverberation with the skull. Resolving power is degraded in azimuth because the varying attenuation at different parts of the skull will cause uniform point reflectors sometimes to return strong echoes, which will be displayed as a large image.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.