The present study focuses on the design and finite element analysis of a novel hip prosthesis comprising truss structures intended to equally distribute the stress along the prosthesis. The prosthesis is designed to suit three different inclination angles of 20°, 40° and 60°. The three designs were subjected to loading conditions that emulate walking, jogging, and cycling activities via finite element simulations. Using the von Mises yield criterion and Goodman relation for fatigue, it is found that the safety factor and the life cycle of the proposed prosthesis design was within the acceptable range. Furthermore, the truss structures incorporated in the prosthesis appear to reduce the stress levels in the femur.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.