A chemiluminescent immunoassay for theophylline in serum or plasma was developed for the Ciba Corning ACS: 180 automated analyzer. The assay has a limit of quantitation of 0.2 microgram/ml, with a range up to 40 micrograms/ml. The cross-reactivity to metabolites 1,3-dimethyluric acid 3-methylxanthine was 3.4% and 2.5%, respectively. Overall means of 101.0% and 97.8% were determined from dilution linearity and addition studies, respectively. When compared to a high-performance liquid chromatography (HPLC) method, a linear regression of ACS Theophylline = 0.996 (HPLC) + 0.35, r = 0.991, n = 93, was obtained. Similar results were obtained when the ACS assay was compared to other immunoassays. Precision (within-run and total coefficient of variation or CV) was < 6% in the therapeutic range (10-20 micrograms/ml). The performance data demonstrate that the ACS Theophylline assay provides an additional choice for the clinical measurement of the drug.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.