Acetaminophen has been one of the most serious electrochemical interferences to oxidase-based amperometric biosensors that measure H2O2. A study was carried out to investigate various polymer materials for their selectivity as the sensor inner membrane. A composite membrane of cellulose acetate and Nafion was found to eliminate acetaminophen and other electrochemical interferences effectively while at the same time maintaining reasonable diffusivity for hydrogen peroxide. The excellent in vivo performance of the sensor was attributed not only to significantly reduced steady-state sensitivity to acetaminophen but also to very slow acetaminophen response. These features, combined with rapid acetaminophen clearance pharmacokinetics, led to the decreased response as demonstrated in the rat.
A miniaturized amperometric, enzymatic, glucose sensor (outer diameter 0.45 mm) was evaluated after implantation in the subcutaneous tissue of normal rats. A simple experimental procedure was designed for the long-term assessment of the sensor's function which was performed by recording the current during an intraperitoneal glucose load. The sensor was calibrated by accounting for the increase in the current during the concomitant increase in plasma glucose concentration, determined in blood sampled at the tail vein. This made it possible to estimate the glucose concentration in subcutaneous tissue. During the glucose load, the change in subcutaneous glucose concentration followed that in blood with a lag time consistently shorter than 5 min. The estimations of subcutaneous glucose concentration during these tests were compared to the concomitant plasma glucose concentrations by using a grid analysis. Three days after implantation (n = 6 experiments), 79 estimations were considered accurate, except for five which were in the acceptable zone. Ten days after implantation (n = 5 experiments), 101 estimations were accurate, except for one value, which was still acceptable. The sensitivity was around 0.5 nA.mmol-1.l-1 on day 3 and day 10. A longitudinal study on seven sensors tested on different days demonstrated a relative stability of the sensor's sensitivity. Finally, histological examination of the zone around the implantation site revealed a fibrotic reaction containing neocapillaries, which could explain the fast response of the sensor to glucose observed in vivo, even on day 10. We conclude that this miniaturized glucose sensor, whose size makes it easily implanted, works for at least ten days after implantation into rat subcutaneous tissue.
We have developed a miniaturized glucose sensor which has been shown previously to function adequately when implanted in the subcutaneous tissue of rats and dogs. Following a glucose load, the sensor output increases, making it possible to calculate a sensitivity coefficient to glucose in vivo, and an extrapolated background current in the absence of glucose. These parameters are used for estimating at any time the apparent subcutaneous glucose concentration from the current. In the previous studies, this calibration was performed a posteriori, on the basis of the retrospective analysis of the changes in blood glucose and in the current generated by the sensor. However, for clinical application of the system, an on line estimation of glucose concentration would be necessary. Thus, this study was undertaken in order to assess the possibility of calibrating the sensor in real time, using a novel calibration procedure and a monitoring unit which was specifically designed for this purpose. This electronic device is able to measure, to filter and to store the current. During an oral glucose challenge, when a stable current is reached, it is possible to feed the unit with two different values of blood glucose and their corresponding times. The unit calculates the in vivo parameters, transforms every single value of current into an estimation of the glucose concentration, and then displays this estimation. In this study, 11 sensors were investigated of which two did not respond to glucose. In the other nine trials, the volunteers were asked to record every 30 s what appeared on the display during the secondary decrease in blood glucose.(ABSTRACT TRUNCATED AT 250 WORDS)
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.