Modern commercial direct-current current transformers (DCCT) can measure currents up to the kA range with accuracies better than 1 × 10 −5 . We discuss here a DCCT calibration method and its implementation with commercial instruments typically employed in low resistance calibration laboratories. The primary current ranges up to 2 kA; in the current range below 100 A the calibration uncertainty is better than 3 × 10 −7 . An example of calibration of a high-performance DCCT specified for primary currents measurement up to 900 A is discussed in detail.
INTRODUCTIONDirect-current current transformers (DCCT) are the most accurate dc high-current sensors commercially available [1], reaching specified relative accuracies in the 10 −5 range and integral nonlinearities below 10 −6 . The verification of such high performances and the calibration of the DCCT ratio require metrological facilities capable of handling high currents, with high accuracy and automated operability [2][3][4][5]. Ultimate current ratio accuracy is achieved in cryogenic current comparators (CCC) [6]. In a CCC, ratio accuracy is obtained by constraining the magnetic flux (generated by the current being compared) within superconducting shields. An extremely high sensitivity is achieved with a superconducting quantum interference device (SQUID) flux sensor. Even though CCCs capable of handling currents up to 100 A have been realized [7], these devices are research instruments not available in calibration laboratories. Ferromagnetic-core, room-temperature current comparators (CC) are current ratio devices which can achieve ratio errors lower than 10 −7 [8], and can be self-calibrated through step-up procedures [9, 10] with similar levels of uncertainty. Thus, a CC can be employed as current ratio standard in a DCCT calibration setup. Although complex and expensive instruments, high-current CC are common in electrical calibration laboratories, since they are part of commercial resistance ratio bridges employed for the measurement of low-value resistors. These instruments include also current sources, detectors, and firmware for automated operation. The calibration of the DCCT ratio with a reference current ratio standard (possibly having a different nominal ratio) can be performed by different methods. Recent papers [11,12] describe a method based on the comparison of the voltages developed by the secondary currents of the devices being compared on calibrated resistance standards. Here we present a simple method that allows the calibration of the ratio of a DCCT by using commercial components, originally designed for the calibration of low-value resistors. This method does not require calibrated resistance standards; the accuracy, dependent on the primary current, is better than 3 × 10
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.