The new M4KDK Diamentor (PTW-Freiburg) for measuring the air-kerma-area product (KAP) and air kerma (AK) simultaneously was evaluated. The special design of the Diamentor chamber influences the variation of the calibration factor with respect to the field size somewhat differently to what is the case with the classic Diamentor. It is therefore recommended that calibration be performed for a larger field size. If the M4KDK Diamentor is calibrated for a field size of 3.8 cm x 3.8 cm (at the chamber), an uncertainty in the AK value of better than 15% for a field size larger than the calibration field, can be expected. The influence of backscattered radiation on the calibration factor is equivalent to that for the classic Diamentor chamber. The variation in response in accordance with tube potential, is also comparable with that of the classic Diamentor. Dramatic changes in response have to be expected if additional filters are introduced in close proximity to the Diamentor chamber.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.