BACKGROUND: Cell salvage is an essential element in the concept of blood management. Modern devices provide different bowl sizes and sensor-directed programs to optimally adjust to varying clinical situations.
STUDY DESIGN AND METHODS:In an experimental performance study, the discontinuous autotransfusion device XTRA (LivaNova/Sorin) was evaluated using fresh donor blood anticoagulated with heparin 5 U/mL and adjusted to a hematocrit of 10% or 25%, representing orthopedic or cardiac surgery. Test blood was processed with the autotransfusion device XTRA in four different bowls (55 mL, 125 mL, 175 mL, and 225 mL) and in three different program modes (a standard program, an optimized program, and an emergency program).
RESULTS:Processing speed increased with bowl size and with the emergency program (range, 6.4-29.8 mL red blood cells [RBCs]/min). The RBC recovery rate exceeded 90% for all bowls and programs except the 55-mL bowl with the emergency program. Plasma elimination exceeded 95% for all bowls and programs except the 225-mL bowl with the emergency and standard programs. Maximal RBC recovery (range, 94.7%-97.6%) and plasma elimination (range, 98.7%-99.5%) were obtained with the medium-sized bowls (125 mL and 175 mL) and the optimized program. Elimination rates for potassium or plasma free hemoglobin were consistently lower than for protein or albumin and were highest for heparin.
CONCLUSIONS:Increased hematocrit and RBC recovery rates are obtained with the optimized program Popt with the discontinuous autotransfusion device. The emergency program Pem speeds up the process but leads to RBC loss and reduced plasma elimination rates; therefore, it should be restricted to emergency situations. All four different sized bowls have high performance.Plasma elimination is represented best by protein or albumin elimination rates.ABBREVIATIONS: Pem 5 emergency program; PFH 5 plasma free hemoglobin; Popt 5 optimized program; Pstd 5 standard program.From the
The autotransfusion device XTRA was capable of high-quality perioperative PRP preparation, and the bowl size was found to have an influence on the efficiency of the preparation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.