Fluoro-18-deoxyglucose positron emission tomography computed tomography (FDG-PET/CT) and magnetic resonance imaging (MRI), including unenhanced single-shot spin-echo echo planar imaging (SS SE-EPI) and small paramagnetic iron oxide (SPIO) enhancement, were compared prospectively for detecting colorectal liver metastases. Twenty-four consecutive patients suspected for metastases underwent MRI and FDG-PET/CT. Fourteen patients (58%) had previously received chemotherapy, including seven patients whose chemotherapy was still continuing to within 1 month of the PET/CT study. The mean interval between PET/CT and MRI was 10.2+/-5.2 days. Histopathology (n=18) or follow-up imaging (n=6) were used as reference. Seventy-seven metastases were detected. In nine patients, MRI and PET/CT gave concordant results. Sensitivities for unenhanced SS SE-EPI, MRI without SS SE-EPI and FDG-PET/CT were, respectively, 100% (p=9 x 10(-10) vs PET, p=8 x 10(-3) vs MRI without SS SE-EPI), 90% (p=2 x 10(-7) vs PET) and 60%. PET/CT sensitivity dropped significantly with decreasing size, from 100% in lesions larger than 20 mm (identical to MRI), over 54% in lesions between 10 and 20 mm (p=3 x 10(5) versus unenhanced SS SE-EPI), to 32% in lesions under 10 mm (p=6 x 10(-5) versus unenhanced SS SE-EPI). Positive predictive value of PET was 100% (identical to MRI). MRI, particularly unenhanced SS SE-EPI, has good sensitivity and positive predictive value for detecting liver metastases from colorectal carcinoma. Its sensitivity is better than that of FDG-PET/CT, especially for small lesions.
Background
This paper describes the National Electrical Manufacturers Association (NEMA) system performance of the Discovery MI 3-ring PET/CT (GE Healthcare) installed in Bruges, Belgium. This time-of-flight (TOF) PET camera is based on silicon photomultipliers instead of photomultiplier tubes.
Methods
The NEMA NU2-2012 standard was used to evaluate spatial resolution, sensitivity, image quality (IQ) and count rate curves of the system. Timing and energy resolution were determined.
Results
Full width at half maximum (FWHM) of spatial resolution in radial, tangential and axial direction was 4.69, 4.08 and 4.68 mm at 1 cm; 5.58, 4.64 and 5.83 mm at 10 cm; and 7.53, 5.08 and 5.47 mm at 20 cm from the centre of the field of view (FOV) for the filtered backprojection reconstruction. For non-TOF ordered subset expectation maximization (OSEM) reconstruction without point spread function (PSF) correction, FWHM was 3.87, 3.69 and 4.15 mm at 1 cm; 4.80, 3.81 and 4.87 mm at 10 cm; and 7.38, 4.16 and 3.98 mm at 20 cm. Sensitivity was 7.258 cps/kBq at the centre of the FOV and 7.117 cps/kBq at 10-cm radial offset. Contrast recovery (CR) using the IQ phantom for the TOF OSEM reconstruction without PSF correction was 47.4, 59.3, 67.0 and 77.0% for the 10-, 13-, 17- and 22-mm radioactive spheres and 82.5 and 85.1% for the 28- and 37-mm non-radioactive spheres. Background variability (BV) was 16.4, 12.1, 9.1, 6.6, 5.1 and 3.8% for the 10-, 13-, 17-, 22-, 28- and 37-mm spheres. Lung error was 8.5%. Peak noise equivalent count rate (NECR) was 102.3 kcps at 23.0 kBq/ml with a scatter fraction of 41.2%. Maximum accuracy error was 3.88%. Coincidence timing resolution was 375.6 ps FWHM. Energy resolution was 9.3% FWHM. Q.Clear reconstruction significantly improved CR and reduced BV compared with OSEM.
Conclusion
System sensitivity and NECR are lower and IQ phantom’s BV is higher compared with larger axial FOV (AFOV) scanners like the 4-ring discovery MI, as expected from the smaller solid angle of the 3-ring system. The other NEMA performance parameters are all comparable with those of the larger AFOV scanners.
Use of digital tomosynthesis of the abdomen results in improved detection of urinary stones in general over digital radiography with only a slight increase in effective dose. Use of tomosynthesis, however, was not associated with major improvement in the diagnosis of ureteral stones. The technique has potential as an alternative imaging technique in the detection and follow-up of urinary stones.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.