We describe basic principles of wavelength-tuning interferometry and demonstrate its application in ophthalmology. The advantage of this technique compared with conventional low-coherence interferometry ranging is the simultaneous measurement of the object structure without the need for a moving reference mirror. Shifting the wavelength of an external-cavity tunable laser diode causes intensity oscillations in the interference pattern of light beams remitted from the intraocular structure. A Fourier transform of the corresponding wave-number-dependent photodetector signal yields the distribution of the scattering potential along the light beam illuminating the eye. We use an external interferometer to linearize the wave-number axis. We obtain high resolution in a model eye by slow tuning over a wide wavelength range. With lower resolution we demonstrate the simultaneous measurement of anterior segment length, vitreous chamber depth, and axial eye length in human eyes in vivo with data-acquisition times in the millisecond range.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.