Controlled changes in the Cs+ primary ion beam density and analytical expressions describing the sources (bulk concentration, memory effect, and adsorption of N from the residual vacuum) of secondary ion mass spectrometry analyte secondary ion intensities were used to determine the contributions to the N secondary ion intensity obtained during the analysis of trace levels of N in bulk SiC. This methodology allows the determination of N concentrations that can be substantially less than the apparent N secondary ion background intensity. It was shown that for the Cameca IMS-6F instrumental conditions used, memory effect is the main contributor to the N background signal. Taking into consideration the Cs+ beam diameter, the raster size, the diameter of the ion-extracted area, and the impurity secondary ion intensity, an optimized combination of primary ion beam current and raster size was determined that resulted in the best detection limit for N in bulk SiC. This detection limit for N in bulk SiC (∼6×1014 atoms/cm3) was obtained for a primary ion current of 100 nA, a raster size of 45 μm×45 μm, and a secondary ion extraction area diameter of 30 μm.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.