A phase fluorometer for the measurement of fluorescence lifetimes was constructed from commercially available components. The instrument was tested by using optical delays of 4 and 2 cm, showing an accuracy of 10 ps. Lifetimes, as short as 0.3 ns, obtained by the quenching of fluorescein by KI, were analyzed with a standard deviation of 3 ps. The lifetime resolving power was checked using mixtures of acridine and quinine–sulphate and least-squares fitting procedures. Accurate amplitude ratios were obtained with the technique of phase-sensitive detection [J. R. Lakowicz and H. Cherek, J. Biochem. Biophys. Methods 5, 19 (1981)].
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.