:In this paper, biomedical Mg-Zn-Ca alloy was processed by high pressure torsion (HPT) up to a maximum of 5 revolutions at room temperature and 7.5GPa. The microstructures of as-received material and HPT material were studied, as well as the corrosion properties and degradation mechanism of the alloy immersed in simulation body fluid (SBF). It revealed that the average grain size of as-received material was 11μm and most of the second phases distributed along the grain boundaries. The corrosion pits formed near the grain boundaries where the corrosion rate was much faster than that of other areas. After HPT processing for 5 revolutions, not only the grain size was significantly refined to 130-150nm, but also the second phases distributed uniformly. These microstructural variations caused the entire surface of the alloy corroded almost at the same corrosion rate. The corrosion interface of the center region and the edge region presented the uniform and gentle profile curve without any obvious corrosion pits. The HPT material in SBF tended to degrade by uniform corrosion mode.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.