Aiming at the method of indoor calibration of a transmission visibility meter has not been established; a new calibration method by measuring the luminous transmittance of attenuator piece was established. Based on the high homogeneity stability lighting and the multi-spectral transmittance analysis technique, a calibration system of the transmission visibility meter by measuring the luminous transmittance of the attenuator piece was designed. The experiment was designed by measuring the luminous transmittance with the range of 0.2-99.2% of a series of calibrated attenuators and the visibility in the range of 40 m to 20 km, then the luminous transmittance errors, the meteorological optical range (MOR) errors and their uncertainty were obtained. The result showed that the luminous transmission error was within ±0.18%, which satisfied the calibration requirement for transmission visibility meter transmittance; the MOR calibration error met the accuracy requirements of the International Civil Aviation Organisation for visibility measurement; this work will have guiding significance in establishing a calibration procedure for the transmission visibility meter.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.