In a recent study, we compared a new intraarterial fiberoptic "optode" probe to continuously measure arterial oxygen and carbon dioxide tensions and pH with intermittently drawn blood samples in patients undergoing surgery. In one patient with a diagnosis of Arnold-Chiari type I malformation with outflow obstruction of the fourth ventricle, a major pulmonary air embolism occurred while the patient was undergoing suboccipital craniectomy and cervical laminectomy in the prone position. Three hours after the incision the optode-displayed oxygen tension decreased from a stable value of 225 +/- 8 mm Hg to 63 mm Hg over a 10-minute period. During the same interval, carbon dioxide tension increased and end-tidal carbon dioxide decreased; shortly thereafter, transcutaneous oxygen tension decreased also. Within 20 minutes after the inspired gas mixture was changed to 100% oxygen, the patient's respiratory variables returned to near baseline. No further complications ensued. This is the first time continuously monitored arterial oxygen tension values during a pulmonary embolism have been reported. With further refinement, intraarterial optode probes will add another valuable method of detecting pulmonary air embolism.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.