The Smart canula concept allows for collapsed cannula insertion, and self-expansion within a vein of the body. (A) Computational fluid dynamics, and (B) bovine experiments (76+/-3.8 kg) were performed for comparative analyses, prior to (C) the first clinical application. For an 18F access, a given flow of 4 l/min (A) resulted in a pressure drop of 49 mmHg for smart cannula versus 140 mmHg for control. The corresponding Reynolds numbers are 680 versus 1170, respectively. (B) For an access of 28F, the maximal flow for smart cannula was 5.8+/-0.5 l/min versus 4.0+/-0.1 l/min for standard (P<0.0001), for 24F 5.5+/-0.6 l/min versus 3.2+/-0.4 l/min (P<0.0001), and for 20F 4.1+/-0.3 l/min versus 1.6+/-0.3 l/min (P<0.0001). The flow obtained with the smart cannula was 270+/-45% (20F), 172+/-26% (24F), and 134+/-13% (28F) of standard (one-way ANOVA, P=0.014). (C) First clinical application (1.42 m2) with a smart cannula showed 3.55 l/min (100% predicted) without additional fluids. All three assessment steps confirm the superior performance of the smart cannula design.
Devices for venous cannulation have seen significant progress over time: the original, rigid steel cannulas have evolved toward flexible plastic cannulas with wire support that prevents kinking, very thin walled wire wound cannulas allowing for percutaneous application, and all sorts of combinations. In contrast to all these rectilinear venous cannula designs, which present the same cross-sectional area over their entire intravascular path, the smartcanula concept of "collapsed insertion and expansion in situ" is the logical next step for venous access. Automatically adjusting cross-sectional area up to a pre-determined diameter or the vessel lumen provides optimal flow and ease of use for both, insertion and removal. Smartcanula performance was assessed in a small series of patients (76 +/- 17 kg) undergoing redo procedures. The calculated target pump flow (2.4 L/min/m2) was 4.42 +/- 61 L/ min. Mean pump flow achieved during cardiopulmonary bypass was 4.84 +/- 87 L/min or 110% of the target. Reduced atrial chatter, kink resistance in situ, and improved blood drainage despite smaller access orifice size, are the most striking advantages of this new device. The benefits of smart cannulation are obvious in remote cannulation for limited access cardiac surgery, but there are many other cannula applications where space is an issue, and that is where smart cannulation is most effective.
Peripheral cannulation for cardiopulmonary bypass (CPB) is of prime interest in minimally invasive open heart surgery. As CPB is initiated with percutaneous cannulae, venous drainage is impeded due to smaller vessel and cannula size. A new cannula was developed which can change shape in situ and therefore may improve venous drainage. An in vitro circuit was set-up with a penrose latex tubing placed between the preload reservoir and the cannula, encasing the cannula's inlet and simulating the vena cava. The preload (P) was stabilised at 2 and at 5 mmHg respectively. The maximum flow rate was determined for 4 conditions: passive venous drainage (PVD) and assisted venous drainage (AVD) using a centrifugal pump at the 2 preload settings. We compared the results of the prototype cannula to classical femoral venous cannulae: basket 28Fr, a thoracic 28Fr and a percutaneous 27Fr. Under PVD conditions and a CVP of 2 mmHg, the prototype cannula's flow rate outperformed the next best cannula by 14% (p=0.0002) and 13% under AVD conditions (p=0.0001). Under PVD conditions and a CVP of 5 mmHg, the prototype cannula outperformed the percutaneous cannula by 19% (p=0.0001) and 14% under AVD conditions (p=0.0002). The new cannula outperforms the classical percutaneous venous cannulae during all of the four conditions tested in vitro.
Peripheral access cardiopulmonary bypass (CPB) is initiated with percutaneous cannulae (CTRL) and venous drainage is often impeded due to smaller vessel and cannula size. A new cannula (Smartcanula, SC) was developed which can change shape in situ and, therefore, may improve venous drainage. Its performance was evaluated using a 2-D computational fluid dynamics (CFD) model. The Navier-Stokes equations could be simplified due to the fact that we use a steady state and a 2-dimensional system while the equation of continuity (p constant) was also simplified. We compared the results of the SC to the CTRL using CFDRC (Version 6.6, CFDRC research corporation, Huntsville, USA) at two preloads (300 and 700 Pa). The SC's mass flow rate outperformed the CTRL by 12.1% and 12.2% at a pressures of 300 and 700 Pa, respectively. At 700 Pa, a pressure gradient of 50% was measured for the CTRL and 11% for the SC. The mean velocity at the 700 Pa for the CTRL was 1.0 m.s(-1) at exit while the SC showed an exit velocity of 1.3 m.s(-1). Shear rates inside the cannulae were similar between the two cannulae. In conclusion, the prototype shows greater mass flow rates compared to the classic cannula; thus, it is more efficient. This is also advocated by a better pressure gradient and higher average velocities. By reducing cannula-tip surface area or increasing hole surface area, greater flow rates are achieved.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.