In ionization chamber dosimetry close to brachytherapy sources, the non-uniform photon fluence in the vicinity of the air cavity of the ionization chamber must be corrected for. In this paper, experimentally determined relative non-uniformity correction factors are determined for a PTW23331 chamber and a number of Farmer-type chambers. It is concluded that the 'anisotropic' theory of Bielajew agrees better with the experimental results than the 'isotropic' theory of Kondo and Randolph. The experiments show that neither the material choice nor the radius of the central electrode have any significant effect on the non-uniformity correction factors. Experiments on the wall-material dependence in the non-uniformity correction factor, which is predicted by the anisotropic theory, were inconclusive. Parameters for the Farmer-type chambers for calculation of theoretical non-uniformity correction factors are given. Data to derive anisotropic fluence non-uniformity correction factors are given in tabular form. An error in the Kondo and Randolph equations is reported and the corrected equations presented.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.