A comparison of capacitance at 10 pF was conducted between thirteen participating laboratories from the Asia-Pacific region. Measurements were made between 2004 and 2006. The behaviour of the travelling artefact was consistent with a steady linear drift at a rate of approximately 0.1 µF/F per year. Despite the wide range of capabilities within the region, the results showed good agreement between all but one of the participating laboratories.Main text.
To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
A comparison of capacitance at 100 pF was conducted between thirteen participating laboratories from the Asia-Pacific region. Measurements were made over the period 2004 to 2006. The behaviour of the travelling artefact was consistent with a steady linear drift at a rate of less than 0.1 µF/F per year. Despite the wide range of capabilities within the region, the results showed good agreement between all participating laboratories.Main text.
To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the APMP, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Currently, three single junction-type Thermal Voltage Converter (TVC) standard units represent the highest standard of AC (Alternating Current) voltages owned by the Electrical Metrology Laboratory, Research Centre for Metrology-Indonesian Institute of Sciences. The accuracy of the single junction-type TVC is maintained regularly via intercomparison processes using a one-step build-up and build-down method. To reduce the calibration process quantity, three steps of build-up and build-down measurements that refer to the 4 V measurement point of a HOLT production single junction-type TVC were carried out. The dissemination processes with the best measurement accuracy up to 20 ppm were successfully obtained from measurement points between 1 V and 20 V via 4-1V, 4-2V, 4-3V, 4-6V, 4-10V, and 4-20V formations.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.