Two versions of a new high-resolution x-ray computed tomography system are being developed to screen mutagenized mice in the Oak Ridge National Laboratory Mammalian Genetics Research Facility. The first prototype employs a single-pixel cadmium zinc telluride detector with a pinhole collimator operating in pulse counting mode. The second version employs a phosphor screedCCD detector operating in current mode. The major system hardware includes a low-energy x-ray tube, two linear translation stages and a rotational stage. For the single-pixel detector, image resolution is determined by the step size of the detector stage; preliminary images have been acquired at 100 pm and 250 pm resolutions. The resolution of the phosphor screen detector IS determined by me modulation transrer function of the phosphor screen; images with resolutions approaching 50 pm have been acquired. The system performance with the two detectors is described and recent images are presented.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.