To compare the diagnostic accuracy of conventional film radiography, charge coupled device (CCD) and photostimulable phosphor plate (PSP) digital images and limited cone-beam computed tomography in detecting simulated horizontal root fracture. Root fractures were created in the horizontal plane in 18 teeth by a mechanical force and fragments were relocated. Another 18 intact teeth with no horizontal root fracture served as a control group. Thirty-six teeth were placed in the respective empty maxillary anterior sockets of a human dry skull in groups three by three. Intraoral radiographs were obtained in three different vertical views by utilizing Eastman Kodak E-speed film, CCD sensor, RVG 5.0 Trophy and a PSP sensor Digora, Optime. Cone beam CT images were taken with a unit (3D Accuitomo; J Morita MFG. Corp, Kyoto, Japan). Three dental radiologists separately examined the intraoral film, PSP, CCD and cone beam CT images for the presence of horizontal root fracture. Specificity and sensitivity for each radiographic technique were calculated. Kappa statistics was used for assessing the agreement between observers. Chi-square statistics was used to determine whether there were differences between the systems. Results were considered significant at P < 0.05. Cone beam CT images revealed significantly higher sensitivities (P < 0.05) than the intraoral systems between which no significant differences were found. Specificities did not show any statistically significant differences between any of the four systems. The kappa values for inter-observer agreement between observers (four pairs) ranged between 0.82-0.90 for the 3DX evaluations and between 0.63-0.71 for the different types of intraoral images. Limited cone beam CT, outperformed the two-dimensional intraoral, conventional as well as digital, radiographic methods in detecting simulated horizontal root fracture.
CBCT is a new imaging technique with a considerably lower radiation dose than conventional MSCT. Our study indicates that CBCT is suitable and, in many ways, equivalent to MSCT, for temporal bone imaging in otosclerosis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.