Purpose: This study evaluated the contrast-to-noise ratio (CNR), spatial resolution, and subjective quality of dual imaging plates (DIP) intraoral radiography.
Methods:The DIP and conventional single IP (CSIP) methods both used YCR DT-1 imaging plates (Yoshida Co.). The DIP, comprising a front IP (FIP) and back IP (BIP), was constructed. DIP images were synthesized from the FIP and BIP images. An aluminum step phantom was used to measure the CNR. A line pair gauge was used to measure the spatial imaging resolution. A phantom comprising a porcine mandible embedded in acrylic resin was used for subjective evaluation.
Results:The CNR of the DIP image was 32% higher than that of the FIP image. The spatial resolution achieved using the FIP, DIP, and CSIP was highly comparable except above 4 line pairs/mm, where that of the CSIP was highest. In subjective evaluation, the noise in the DIP images was significantly lower than in those obtained using the FIP and CSIP.
Conclusion:The CNR of the DIP was higher than that of the FIP. The decrease in spatial resolution of the DIP was limited. The subjective image quality of the DIP was higher than that of the FIP.
Purpose:The dual imaging plate (DIP) method, which synthesizes intraoral radiographs from a front imaging plate (FIP) and a back imaging plate (BIP), produces adequate image quality and allows the radiation dose to be reduced. However, there are slight errors in superimposition and alignment between the FIP and BIP. The aim of this study was to establish positional correction in the DIP method and evaluate the effect. Methods: Six sets of two imaging plates were used for imaging a mesh plate and a porcine mandible phantom. Subtraction images between FIP and BIP images were synthesized in four steps: correcting horizontal and vertical direction, rotation, enlargement ratio, and enlargement ratio into 12 blocks. Variance of the pixel value on the subtraction images at each step was compared to evaluate the alignment of FIP and BIP images. Results: The variance of the pixel values in the subtraction images was gradually and significantly decreased by each step of image processing (P < 0.01), indicating that the degree of alignment of FIP and BIP images improved during the image processing.
Conclusion:The present study revealed that it is possible to synthesize more precise DIP images using an additional four-step image processing technique.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.