A simplified QuEChERS method coupled with a small injection volume was developed for the simultaneous determination of 28 pesticides in 6 matrices (apples, cucumbers, tomatoes, luffa, cabbages, and eggplants) using LC-MS/MS. In the sample procedure, acetonitrile with 0.1% (v/v) formic acid was used to extract and no depurative powder was used. Three fortified levels (10, 50, and 100 mg kg À1 ) were determined and the recoveries of 168 analyte/matrix combinations were in the range of 60-120% except for cyromazin, pendimethalin, and fenpropathrin. Half of the 168 LOQs were below 0.1 mg kg À1 , and 29 LOQs were above 1 mg kg À1 . Moreover, four relationships between signal suppression and injection volume were observed ranging from 0.5 mL to 15 mL. For many analyte/matrix combinations, the matrix effects could be reduced to less than 20% if the injection volume was less than a critical value (named critical volume).The critical volume depending on the initial extent of matrix effects was explored and the conclusion was: for weak or medium MEs, usually #2 mL injection volume was needed and for several weak MEs, injection volume # 5 mL can reduce the matrix effect to a negligible level.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.