For the sufficient exclusion of a fracture the dose can be reduced to at least 39%, for the sufficient assessment of bony union and possible inflammatory changes to at least 42%, to position check and foreign body search to at least 23%. By DLR the following question-referred mAs-product at 50 kV are stated: fracture exclusion 2 mAs, bony union, arthritis and osteomyelitis 2.2 mAs, position check and foreign body search 1.25 mAs.
For the sufficient assessment of a possible fracture/of paranasal sinuses/of measurement of the skull the dose can be reduced to at least 56% (phi 31%; SD 14.9%)/40% (phi 27%; SD 9.3%)/18% (phi 14%; SD 4.4%). Digital radiography allows question-referred exposure parameters with clearly reduced dose, so e.g. for fracture exclusion 73 kV/12.5 mAs and to skull measurement 73 kV/4 mAs.
For the sufficient exclusion of a fracture the dose can be reduced on pelvis/LS to at least 39 %/28 % ( 23 %/17 %), for the assessment of sacroiliitis/osteolyses/demineralisation to at least 78 %/44 % ( 34 to 36 %/22 to 26 %), to position check after fracture to at least 15 %/25 % ( 9 %/10 %). Question-related tube current time products of the pelvis and LS can be deduced for various objects.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.