In radiotherapy, point doses via surface dose measurement are confirmed using a dosimeter attached to the patient's skin. This results in an error margin of about 11.8% or higher because the attachment site is visually confirmed. Thus, a digital flexible array dosimeter that can analyse body surface area is needed. Here, a lead (II) iodide (PbI 2 )-based unit cell flexible dosimeter was developed, and its electrical properties were evaluated based on its mechanical flexibility by assessing its reproducibility and linearity for bending. The relative standard deviation (RSD) for measurement by the sensor over 10,000 cycles was 1.74% and 2.21% at 6 MV and 15 MV, respectively, which is higher than RSD's evaluation criterion (<1.5%). Although the linearity of measurement by the sensor over 10,000 cycles met the evaluation criteria of a coefficient of determination R 2 > 0.9990, the sensitivity at 15 MV was lower than that at 6 MV. The results of this study confirm the applicability and usefulness of the PbI 2 -based flexible dosimeter in radiation measurements. K: Dosimetry concepts and apparatus; Materials for solid-state detectors; Radiation damage to detector materials (solid state); Radiotherapy concepts 1Corresponding author.
This study aims to quantitatively evaluate the light and radiation fields of a linear accelerator by developing an optical sensor quality assurance (QA) system using a cadmium sulfide (CdS) photodetector cell and a hardware base frame manufactured in-house. The CdS photodetector cell used as the optical sensor has a size of 3.4 × 4.1 mm2, can measure up to 100 lux of light, and has a response time of approximately 20 to 30 ms. To check the alignment of the light field, it was manufactured with a CdS cell optical fiber holder, and the microcontroller board and control module were configured to scan the light field profile. Additionally, for accurate operation of the sensor, it was programmed using the integrated development environment (IDE) sketch. The alignment check of the radiation field was analyzed using the image data measured with electronic portal imaging device (EPID) using the manufactured hardware base frame. An analysis program written in MATLAB was used to evaluate the alignment of the radiation field. For symmetric fields measured using the optical sensor QA system, the full width at half maximum (FWHM) values of crossline (X) and inline (Y) profiles were 100.391 mm and 98.725 mm, respectively, for a field size of 100 × 100 mm2, and 199.093 mm and 198.886 mm, respectively, for a field size of 200 × 200 mm2. The X and Y FWHM values measured by the sensor system for the light fields were within 1 mm of the field size of the visual inspection. In addition, the FWHM value of the asymmetric radiation field measured using the fabricated frame and EPID was measured from the center bearing ball of the field to the bearing ball in each field of the four directions in the EPID image, and when the X1, X2, Y1, and Y2 jaws were opened by 50 mm, FWHM values were 50.120 mm, 50.240 mm, 48.860 mm, and 49.870 mm, respectively. The corresponding values were 99.520 mm, 99.570 mm, 99.010 mm, and 99.880 mm when the jaw fields were opened by 100 mm. The optical sensor system and hardware base frame developed in this study demonstrated the ability to evaluate both light and radiation fields more simply and quantitatively when compared to the conventional QA process. This QA system will provide medical physicists with more reliable QA results than the conventional QA method that uses graph paper and film.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.