Aqueous ethanol wet-bath simulator solutions are used to perform calibration adjustments, calibration checks, proficiency testing, and inspection of breath alcohol instruments. The Toxicology Bureau of the New Mexico Department of Health has conducted a study to estimate a measurement of uncertainty for the preparation and testing of these wet-bath simulator solutions. The measurand is identified as the mass concentration of ethanol (g/100 mL) determined through dual capillary column headspace gas chromatography with flame ionization detector analysis. Three groups were used in the estimation of the aqueous ethanol wet-bath simulator solutions uncertainty: GC calibration adjustment, GC analytical, and certified reference material. The standard uncertainties for these uncertainty sources were combined using the method of root-sum-squares to give uc = 0.8598%. The combined standard uncertainty was expanded to U = 1.7% to reflect a confidence level of 95% using a coverage factor of 2. This estimation applies to all aqueous ethanol wet-bath simulator solution concentrations produced by this laboratory.
Reporting a measurement of uncertainty helps to determine the limitations of the method of analysis and aids in laboratory accreditation. This laboratory has conducted a study to estimate a reasonable uncertainty for the mass concentration of vaporous ethanol, in g/210 L, by the Intoxilyzer(®) 8000 breath analyzer. The uncertainty sources used were: gas chromatograph (GC) calibration adjustment, GC analytical, certified reference material, Intoxilyzer(®) 8000 calibration adjustment and Intoxilyzer(®) 8000 analytical. Standard uncertainties attributed to these sources were calculated and separated into proportional and constant standard uncertainties. Both the combined proportional and the constant standard uncertainties were further combined to an expanded uncertainty as both a percentage and an unit. To prevent any under reporting of the expanded uncertainty, 0.10 g/210 L was chosen as the defining point for expressing the expanded uncertainty. For the Intoxilyzer(®) 8000, all vaporous ethanol results at or above 0.10 g/210 L, the expanded uncertainty will be reported as ±3.6% at a confidence level of 95% (k = 2); for vaporous ethanol results below 0.10 g/210 L, the expanded uncertainty will be reported as ±0.0036 g/210 L at a confidence level of 95% (k = 2).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.