Multiprogrammable pacemakers have long been subject to inappropriate reprogramming and electromagnetic interference (EMI). A limited clinical experience with DDD pacing systems precludes the significance of such phenomena in these units. Since August 1981, in a series of 140 DDD systems, certain units demonstrated consistent and reproducible resetting to the back-up modes caused by electrocautery. One unit was permanently reset to the VOO mode. These observations suggest the need for renewed caution at the time of surgery; they also underscore the importance of a thorough understanding of any DDD system and careful follow-up so that such phenomena will not be misinterpreted as pacemaker failure, resulting in erroneous pacemaker removal.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.