When an electrical current with a low frequency is applied to a cell, the current passes through the outside of the cell. Thus, impedance measurements at low frequencies cannot be used to determine the pathological change of the cellular organelle taking place inside the cell. However, increasing the frequency of the electrical current makes the capacitive impedance of the cell decrease, allowing the electrical current to flow through the cell. This study presents the design and fabrication of a microfluidic device integrated with a coplanar waveguide open-ended micro-electro-mechanical-systems (MEMS) probe for the impedance measurement of the single HeLa cell in frequencies between 1 MHz and 1 GHz. The device includes a poly-dimethlysiloxane (PDMS) cover with a microchannel and microstructures to capture the single HeLa cell and a conductor-backed CPW fabricated using a silicon chip and two printed circuit boards (PCB). The effects of the substrate on the characteristic impedance of the conductor-backed coplanar waveguide (CBCPW) structure were investigated under three conditions by utilizing a time-domain reflectometer (TDR). Finally, impedance measurements using the proposed device and a vector network analyzer (VNA) are demonstrated for de-ionized (DI) water, alcohol, PBS, and a single HeLa cell.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.