During a 6-month study we critically evaluated the accuracy of the AutoMicrobic system Gram-Negative Identification Card (Vitek Systems, Inc., Hazelwood, Mo.) in identifying glucose-nonfermenting gram-negative bacilli by testing 419 selected isolates in parallel with a conventional reference method. Of 356 isolates included in the AutoMicrobic system profile, a total of 307 (86.2%) were correctly identified, 36 (10.1%) were not identified, and 13 (3.7%) were misidentified. Fifty-eight of 63 (92%) isolates not included in the profile were correctly reported as "unidentified organisms." Overall, if the first-choice identification was always accepted, only 18 (4.3%) isolates would have been incorrectly reported. When first-choice identifications appended with the special message "questionable biopattern" were rejected, and organisms were screened for characteristic odor and antimicrobial susceptibility before final acceptance of the AutoMicrobic system report, the number of misidentifications was reduced to 5 (1.2%). The average time to identification with the AutoMicrobic system Gram-Negative Identification Card was 15 h. This compares favorably with the 65 h required by the reference method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.